69
SpO 2 and PR alarm limits:
Max. Upper Limit
SpO
100
2
PR
250
The default SpO 2 and PR alarm limits:
Parameters
SpO
Adult
2
Pediatric
Neonatal
PR
Adult
Pediatric
Neonatal
• SWEEP
Available options are 12.5mm/s, 25.0 mm/s.
• AVG TIME
4s, 8s, 16s represent times that SpO
• DEFAULT
Pick this item to access the SpO
DEFAULT CONFIG or USER DEFAULT CONFIG. After selecting one item and exiting the dialog box,
the system will pop up the dialog box asking for the your confirmation.
12.6 Alarm description and Prompt
SpO 2 Alarm Message
When the alarm switches are set to ON in relevant menus, the physiological alarms caused by the parameter
exceeding the alarm limit may possibly trigger the recorder to automatically output alarming parameter value
and corresponding waveforms.
Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring
during SpO
measurement.
2
Physiological alarm:
Message
SpO
TOO HIGH
2
SpO
TOO LOW
2
PR TOO HIGH
PR TOO LOW
Technical alarms:
Message
SpO
SENSOR OFF
2
SpO
INIT ERR
2
SpO
INIT ERR 1
2
SpO
INIT ERR 2
2
SpO
INIT ERR 3
2
SpO
INIT ERR 4
2
SpO
INIT ERR 5
2
SpO
INIT ERR 6
2
SpO
INIT ERR 7
2
SpO
INIT ERR 8
2
Max. Upper Limit
100
100
95
120
160
200
average value is counted.
2
DEFAULT CONFIG dialog box, in which you can select FACTORY
2
Cause
SpO
measuring value is above upper alarm limit.
2
SpO
measuring value is below lower alarm limit.
2
PR measuring value is above upper alarm limit.
PR measuring value is below lower alarm limit.
Cause
SpO
sensor may be di-
2
sconnected from the patient
or the monitor.
SpO
module failure
2
Min. Lower Limit
0
30
Alarm Level
Remedy
LOW
Make sure that the monitor and
the patient are in correct connec-
tion with the cables.
HIGH
Stop using the measuring fun-
ction of SpO
medical engineer or our service
staff.
Step
1
1
Min. Lower Limit
90
90
85
50
75
100
Alarm Level
User-selectable
User-selectable
User-selectable
User-selectable
module, notify bio-
2