61
Figure 11-13 ARR Recall Menu
• UP-DOWN Observe other event lists of other page.
• CURSOR
Select the Arr. event, whose name is displayed in a protruding frame.
• RENAME
Rename the selected Arr. event, whose name is displayed in a sunken frame.
Switch the knob until the name you want appears.
• WAVE
To display the Arrhythmia waveform, time and parameter value.
• UP-DOWN To observe waveforms of other Arrhythmia events.
• RECORD
To print out displayed Arrhythmia event.
• EXIT
To return to ARR RECALL menu of Arrhythmia event.
Figure 11-14 ARR Wave Recall Menu
Note
If there are more than 60 Arrhythmia events, the latest will be retained.
ARR ALARM
The alarm is triggered when an Arrhythmia occurs. If the ALM is ON, the alarm sounds and the alarm indica-
tor flashes. If the REC is ON, the alarm record will be printed out (4 seconds prior to and after the alarm, with
the ECG waveforms of analysis channel).