If the maximum value of a parameter is less than or equal to the minimum value, the setting process will not proceed. A beep
will sound and the device will automatically return to the screen for setting the minimum value.
Once the upper threshold of Pulse Rate has been set, confirming with OK, the sequence
will continue with the ON/OFF setting of the following alarm conditions:
Finger not inserted
Sensor unplugged
Battery discharged
The image illustrates an example of setting for an alarm condition: in this case it is the
alarm for "finger is not correctly inserted " .
SpO2 sampling rate
This function allows to set the time that elapses between the sampling/recording of two consecutive oximetry readings; touch one of the
two visualized icons: 2 seconds or 4 seconds, then touch OK to set the selected value and the device will automatically return to the
service menu.
Pulse tone ON/OFF
This setting enables the Pulse rate tone (audible beep) during oximetry testing.
WARNING
The Pulse rate tone (beep) is always disabled during sleep oximetry testing
Default alarms (factory setting)
This function allows to restore all factory default alarms setting. To confirm touch the
"YES" icon. The settings return to the factory values.
The image on the right shows the factory default alarms setting.
After a few seconds, the device will show the oximetry setting menu.
Select predicted
Select the item by touching the display.
A list of predicted values is shown; select the Predicted value desired.
Select the pair to use and touch OK. The Predicted values are set and the device returns to the Service Menu.
Select standard
Select the item by toughing the display.
Select the standard to be used (ATS/ERS, or NHANES III) and touch OK, the setting takes effect and the device returns to the Service
Menu.
WARNING
If the NHANES III standard is selected it is not possible to set or modify the predicted values.
Set parameters
SPIRODOC
Adult
Pediatric
ERS
Knudson
Knudson
Knudson
USA
Knudson
ERS
Zapletal
MC-Barcelona
Zapletal
JRS
JRS
PEREIRA
PEREIRA
CHINESE HK
CHINESE HK
cod. 980156
Rev 4.2.1
spirodoc
User manual
EN 27/48