When the alarm is activated by more than one physiological alarm condition, each parameter will be displayed in reverse mode.
If the alarm is triggered by a technical condition, the relevant warning message is displayed, for example:
Auditory Alarm Indicator
Audible alarms can be heard in a quiet environment. The medium priority audible alarm has a "du-du-du" tone that repeats every 5
seconds. The acoustic alarm signal can be temporarily disabled while an alarm condition is in progress. The duration of the paused audio,
the time interval in which the alarm system or part of the alarm system does not generate an audible alarm signal, is a maximum of 2
minutes.
The sound pressure level of the alarm tone is about 55 dB, in complies with the standard.
Other beeps (acoustic signals):
•
"Beep" with frequency of the cardiac pulse
•
If the oximetry test has been interrupted
switched on again
The specifications for both the oximetry and for the cardiac pulse are the same regardless of which of the above mentioned oximetry
sensors is used.
1.7.4
Other features
Memory
keyboard
Display
Interface
Bluetooth interface
Duration of the 3,7V lithium battery
Power supply
Battery charger
Dimensions
Weight
Type of electrical protection
Degree of electrical protection
Grade of protection against water ingress
Safety level in the presence of inflammable anaesthetic
gas, oxygen or nitrogen
Conditions of use
Storage conditions
Transport condition
Operating conditions
Applied norms
SPIRODOC
WARNING
FINGER inserted incorrectly
due to low battery
an intermittent beeping will be heard for 5 seconds when the device is
Memory capacity for over 10000 spirometric tests
The number depends on the individual configuration, so it cannot be
determined more closely
Absent, display touch screen
Black and White LCD touchscreen with
USB, Bluetooth
frequenzy range = 2402-2480 MHz
rated RF power output = 0.001 W
frequency tolerance = 20 ppm
type of antenna = permanently attached
gain of antenna = 0 max dBi
Approx 500 charge cycles, under normal conditions of use
Battery pack Li-ion 3.7 V 1100mAh
Voltage = 5VDC
Current = 500 mA
Connector = micro USB type B
101x48x16 mm;
turbine housing 46x47x24 mm
Central unit 99g (including batteries)
Turbine housing 17g
Internally powered
BF
IPX1
Device not suitable
Device for continuous use
Temperature: MIN -20 °C, MAX + 60 °C
Humidity :MIN 10% RH; MAX 95%RH
Athmospheric pressure: 50kPa, 106 kPa
Temperature: MIN -20 °C, MAX + 60 °C
Humidity :MIN 10% RH; MAX 95%RH
Athmospheric pressure: 50kPa, 106 kPa
Temperature: MIN + 10 °C, MAX + 40 °C;
Humidity: MIN 10% RH; MAX 95%RH
Athmospheric pressure: 70kPa, 106 kPa
IEC 60601-1:2005 + A1: 2012 (Electrical Safety)
IEC 60601-1-2:2015 (EMC)
ATS/ERS Guidelines: 2005
ISO 26782: 2009
ISO 23747: 2015
EN ISO 14971: 2019
ISO 10993-1: 2018
2011/65/UE Directive
EN ISO 15223-1:2021
cod. 980156
Rev 4.2.1
spirodoc
User manual