CS-3000 RF Generator Unit Operator Manual
Table 1 – List of Troubleshooting Symptoms & Actions
Symptom
No displays or indicators when the
RF Generator is turned on
IMP LIMIT warning, indicating
impedance out of range
Error LED indicator illuminates
and one of the following Fault
messages appears in Message
window:
RF EFF LIMIT
ROM CRC FAILURE
RAM FAILURE
TIMER FAILURE
+48 VOLT SUPPLY FAIL
-12 VOLT SUPPLY FAIL
+12 VOLT SUPPLY FAIL
+5 VOLT SUPPLY FAIL
IMPEDANCE TEST FAIL
RF POWER TOO HIGH WHEN
OFF
RF POWER TOO HIGH WHEN
ON
RESISTIVITY FAILED
MEASURED POWER
CALCULATED POWER
FAN FAULT
GROUND FLOAT ON A/D
CAL VOLTAGE ON A/D
AMBIENT TEMP FAIL
Error LED indicator shows a
button is stuck and one of the
following Fault messages appears:
RF ON BUTTON STUCK
MODE BUTTON STUCK
POWER/TEMP UP BUTTON
STUCK
POWER/TEMP DOWN BUTTON
STUCK
TIME UP BUTTON STUCK
TIME DOWN BUTTON STUCK
RF power does not turn on when
footswitch is depressed
RF interferes with ultrasound and
other equipment
AtriCure, Inc.
Action
• Be sure the generator is plugged into a
working electrical outlet.
• Check power switch on back of
generator.
• Unplug and check the fuse on the rear
panel. Fuses should only be replaced
with LittelFuse 2183.15, Time Delay (Slo
Blo) rated at 3.15A, 250VAC.
• Check connections to coagulation device
and indifferent, dispersive electrode.
• Ensure device is properly placed on
tissue site.
• Check position of indifferent, dispersive
electrode on patient's back or thigh.
• If problem persists, replace the
coagulation device and indifferent,
dispersive electrode.
• Power unit OFF, then Power ON again.
• Allow generator to run through normal
start-up self-diagnostics.
• If generator returns to Error state and the
problem persists, contact your AtriCure,
Inc. representative.
• Depress and release the indicated button
to determine if the button remains stuck.
• Power unit OFF, then Power ON again.
• Allow generator to run through normal
start-up self-diagnostics.
• If generator returns to Error state and the
problem persists, contact your AtriCure,
Inc. representative.
• Verify the footswitch is connected to the
front panel of the generator
• Ensure Time is not set to 0.
• Ensure IMP LIMIT error is not detected -
impedance should be within 30 – 500Ω.
• Ensure a coagulation device is connected
to the generator
• Ensure the indifferent, dispersive
electrode is attached to the patient and
connected to the generator
• Check the footswitch by unplugging its
cable from the generator, placing a
thumb over the foot pedal connector then
depressing the pedal. If the footswitch is
functioning properly, air should be
expelled through the connector as the
footswitch is depressed.
• Ensure the cables from the electrodes do
not cross the cables from the ultrasound
probe or other equipment.
• Changing settings on the ultrasound may
alleviate image interference.
Symptom
Impedance is greater than 500 Ω
at the beginning of RF power
delivery
Impedance is less than 30 Ω at
the beginning of RF power
delivery
LBL-2123 Rev. F
Action
• Ensure the indifferent, dispersive
electrode is properly attached.
• Check all connections.
• Clean off any coagulum from the
coagulation device.
• Ensure device is properly placed on
tissue site.
• Reposition the coagulating device
Page 14/ 94