Installation
EN-
8.3
Configuring the unit in
US
DBSWIN
The configuration is done with VistaConfig, which
is automatically installed during the installation of
DBSWIN or VistaEasy.
Start > All Programs > Air Techniques > Vis-
❯
taEasy > VistaConfig.
Click
.
❯
This updates the list of connected devices.
If the device isn't found, enter the IP address
❯
manually and click Register device manually.
Activate the connected device in the Regis-
❯
tered column.
You can also register multiple devices.
Use the ScanX device configuration window to
change the device name, (name), to manually
enter an IP address and to request information.
Click
.
❯
28
Entering a permanent IP address (recom-
mended)
To reset the network settings, keep the
unit reset button pressed for 15 - 20 sec-
onds while switching on.
Deactivate DHCP.
❯
Enter the IP address, subnet mask and gate-
❯
way.
Click on Apply.
❯
This saves the configuration.
Testing the device
You can scan an X-ray image to check if the unit
is properly connected.
Select the Test tab.
❯
Select the unit from the Registered Devices
❯
selection list.
Select the mode class.
❯
Select the mode.
❯
Click on Scan Image.
❯
For the scanning of an image plate, see "11.3
❯
Scanning the image data via a computer".
8.4
Prepare the X-ray cassette for
an extraoral phosphor stor-
age plate
The existing X-ray cassettes used with X-ray film
can be retrofitted to enable the use of phosphor
storage plates.
Remove the Intensifying screen(s) from the X-
❯
ray cassette.
2151100018L40 2012V016