Check the PSP with a homogeneous test
image
Scratches on the surface of the PSP may be visi-
ble in the X-ray image and may impair the ability
to diagnose the X-ray images. If scratches are
visible in the image that may impair the diagnos-
tics, the PSP must be discarded.
Fig. 10: PSP with scratches
Place the PSP on a level surface at a distance
of approximately 30 cm from the X-ray tube.
Doing this, make sure that the active side of
the PSP faces the X-ray tube.
Set the exposure time and setting values on
the X-ray unit for a molar x-ray image (see "8.5
X-ray unit settings").
Scan the PSP with the PSP scanner at high
resolution.
Check the image for homogeneity.
No scratches visible in the image: PSP can still
be used.
Scratch is visible in the image: discard the PSP.
2151100018L40 2012V016
Check X-ray system
To check the X-ray system, take an X-ray image
with the 2D X-ray test phantom (refer to "3.3
Optional items"). This can be used to check the
image produced with the X-ray system for homo-
geneity, resolution, contrast and artifacts.
Take an X-ray image with the 2D X-ray test
phantom. Comply with the instructions for use
of the test body.
Read the PSP.
Check the image for homogeneity, resolution,
contrast and artifacts.
If errors are visible in the image, contact a serv-
ice technician.