Port
Purpose
2006 TCP
Device data
Event protocol data
1)
514
UDP
2005 TCP,
Diagnosis
23 TCP
The port can vary depending on the con-
1)
figuration.
Configuring WLAN on the unit
If the unit is to be operated via WLAN, the con-
nection needs to be configured on the unit.
In order to establish a secure WLAN con-
nection, we recommend encrypting the
WLAN network with WPA2.
The quality and transmission range of the
WLAN connection can be reduced by
environmental conditions (e.g. thick walls,
other WLAN devices). Mind the signal
strength when you select a suitable loca-
tion for set-up.
Requirements:
ü You need to be logged-in on the unit as
Administrator or Service Technician (Settings >
Access Levels > Administrator/Service Tech-
nician).
Check the WLAN settings with your Network
❯
Administrator.
Tap the following on the touch screen: Settings
❯
> System Settings > Network.
Under Interface, select the WLAN option and
❯
confirm with OK.
Configure the WLAN.
❯
Confirm with OK.
❯
When the unit is first connected to a com-
puter, it applies the language and time
settings of the computer.
8.2
Configuring the unit in
VisionX
Configuration is performed directly in VisionX.
> Select the units.
❯
2151100018L40 2012V016
Mark the connected unit in the list.
❯
Service
Syslog
Telnet,
SSH
Click on Edit connection settings.
❯
The unit name (designation) can be changed
❯
and information queried in General.
An IP address can be entered manually and
❯
DHCP can be activated / deactivated in Con-
nection.
Extended functions e. g. IP address 2 can be
❯
set in Extended.
Entering a permanent IP address (recom-
mended)
To reset the network settings, keep the
unit reset key pressed for 15 - 20 sec-
onds while switching on.
Deactivate DHCP in Connection.
❯
Enter the IP address, subnet mask and gate-
❯
way.
Navigate back to Units via the navigation bar
❯
or close Flyout using
This saves the configuration.
Testing the device
You can scan an X-ray image to check if the unit
is properly connected.
Open VisionX.
❯
Create an X-ray station for the connected unit.
❯
Log-in the demo patient (patient ID:
❯
DEMO0001).
Select the acquisition type (e. g. Intraoral).
❯
To scan a PSP, see "11.3 Scanning the image
❯
data via a computer".
Installation
EN-
US
.