Installation
EN-
Only connect units when it is safe to do so and
❯
when there is no risk of damage or harm to the
US
surroundings.
If it is not completely clear from the data sheet
❯
of the unit that such connections can be safely
made or if you are in any doubt, always get a
suitably qualified person (e.g. the relevant man-
ufacturer) to verify that the setup is safe.
A template for the system manufacturer's
declaration in accordance with Article 12
of Directive 93/42/EEC can be found in
our download section at www.duerrden-
tal.com (document no. 9000-461-264).
Only connect peripheral units (e.g. computer,
❯
monitor, printer) that conform at least to the
requirements set out in the standards,
IEC 60950‑1 (EN 60950‑1) or IEC 62368‑1
(EN 62368‑1).
Connecting the unit via the network cable
Plug the enclosed network cable into the net-
❯
work socket of the device.
26
8
Commissioning and first
start-up
NOTICE
Short circuit due to the build-up of
condensation
Do not switch on the unit until it has
❯
warmed up to room temperature and
is dry.
The unit supports the following imaging pro-
grams:
– VisionX manufactured for Air Techniques
– DBSWIN manufactured for Air Techniques
– Please inquire about third-party software
Always use the current version of the
imaging program in the commissioning of
the device. Check the version of the
enclosed imaging program versus the ver-
sions available at www.airtechni-
ques.com.
8.1
Configuring the network
Switch on the network devices (router, PC,
❯
switch).
Check if TCP port 2006 and UDP port 514 are
❯
enabled in the firewall; enable them, if neces-
sary.
If you are using the Windows firewall, you do
not need to check the ports since you will be
asked whether you want to enable them during
the driver installation process.
Network configuration
Various options are available for network configu-
ration:
ü Automatic configuration via DHCP.
ü Automatic configuration via Auto-IP for direct
connection of device and computer.
ü Manual configuration.
Configure the network settings of the device
❯
using the software or, if applicable, the touch
screen.
Check the firewall and release the ports, if
❯
applicable.
Network protocols and ports
Port
Purpose
45123 UDP,
Device recognition and
45124 UDP
configuration
2151100018L40 2012V016
Service