WARNING
Contamination of the unit
Clean and disinfect the barrier enve-
❯
lope before removing the PSP.
In the event of heavy soiling, e. g. from blood,
❯
dry clean the barrier envelope and protective
gloves, e. g. wipe with a clean cellulose cloth.
Disinfect the barrier envelope and protective
❯
gloves with a disinfectant wipe.
Alternatively, use a spray disinfectant on a soft,
lint-free cloth.
Allow the barrier envelope to fully dry.
❯
Pull off the protective gloves, disinfect and
❯
clean the hands.
2151100018L40 2012V016
NOTICE
Powder from the protective gloves on
the PSP damages the unit during
scanning
Completely clean all traces of the pro-
❯
tective glove powder from your hands
before handling the PSP.
Tear the barrier envelope and place the phos-
❯
phor storage plate in the transfer box.
If the active side of the phosphor storage plate
is facing up, turn the phosphor storage plate
over immediately.
Usage
EN-
US
37