If a scanning job is started via the imaging soft-
ware, the image is automatically transmitted to
the computer.
If a scanning job is started via the touch screen,
the image is saved to the memory card and
needs to be transferred to the computer later on.
After scanning, the phosphor storage plate runs
through the erasure unit. Image data still held on
the phosphor storage plate is erased with the aid
of bright light.
The phosphor storage plate is then ejected for
re-use.
The unit can scan up to four PSPs simultane-
ously at the same resolution.
Operating elements
1
Touch screen
2
On/off switch
The touch screen allows the unit to be operated
when it is not connected to a computer. Instruc-
tions can be entered on the touch screen either
with the tip of a finger or the Stylus.
The Help button can be used to open a help
page for the respective page. The Messages
button can be used to call up current messages.
2151100018L40 2012V016
Connections
The connections are located on the rear of the
unit.
1
2
1 2
3
4
1
Reset button
2
Connection for power supply unit
3
AUX connection for diagnostic units
4
Network connection with status LED
ScanManager
When the ScanManager is enabled, more than
one X-ray job can be transmitted simultaneously
to the unit from different computers. The unit
manages the X-ray jobs in a queue from which
the respective X-ray job can be selected using
the touch screen and then executed.
Without ScanManager, the unit processes one X-
ray job at a time and is blocked until this job has
been completed. During this time, no further X-
ray jobs can be transmitted to the unit from other
computers.
ScanManager can be enabled via Set-
tings > System Settings> Operating
Type.