CAUTION
Loss of image data caused by stray
ambient light entering the unit
Remove the intraoral/extraoral plate
protector only after the phosphor stor-
age plate has dropped into the collect-
ing tray.
Place a finger on the printed surface of the
phosphor storage plate and push downwards
until the phosphor storage plate is drawn in
automatically.
Scanning progress is displayed on the touch
screen. Image data is saved automatically.
The touch screen only shows a preview
as an initial impression of the X-ray image.
Limitations to image previews occur due
to image size and/or exposure conditions.
For purposes of diagnosis, the X-ray
image must be viewed on a diagnostic
monitor.
After it has been scanned, the phosphor storage
plate is erased and drops into the collecting tray.
Save the X-ray image.
Remove the PSP and prepare it for recording
another X-ray image.
Remove the plate protector.
Transmitting image data to the computer
X-ray images created using the touch screen of
the device are saved to a memory card. These X-
2151100018L40 2012V016
ray images can be imported into an imaging soft-
ware via a network connection.
Connect the unit to the network.
Start the imaging software.
Start the image import via the imaging software
(further information can be found in the manual
of the imaging software).
Save the image data.
The image data on the memory card is erased
automatically as soon as the transfer has been
successfully completed.