Ghosting or double exposure
on X-ray image
X-ray image mirrored in one
corner
Shadow on the X-ray image
X-ray image cut off, part miss-
ing
Software unable to combine
the data to make a complete
image
2151100018L40 2012V016
Possible cause
PSP exposed twice
PSP not sufficiently erased
PSP creased during X-ray expo-
sure
PSP removed from the barrier
envelope before scanning
A metal part of the X-ray tube is
in front of the X-ray beam
Faulty edge masking in imaging
software
The X-ray dose on the PSP was
insufficient
Amplification (HV value) is set
too low in the software
Unsuitable scanning mode
selected
The setting for the threshold
value is too high
Troubleshooting
Remedy
❯ Only expose the PSP once.
❯ Check the erasure unit for
proper function.
❯ Inform a service technician, if
the problem persists.
❯ Do not bend the PSP.
❯ Do not handle PSPs without a
barrier envelope.
❯ Store the PSP in a barrier
envelope.
❯ Recording an X-ray image,
make sure there are no metal
parts between the X-ray tube
and the patient.
❯ Check X-ray tube.
❯ Deactivate edge masking.
❯ Increase X-ray dose.
❯ Increase amplification (HV
value).
❯ Select a suitable scanning
mode.
❯ Reduce the threshold value.