Usage
EN-
Inspection inter-
US
val
Daily / before each
use
Monthly
Every 3 months
48
Work
❯ Clean the PSP, if necessary.
❯ Check the PSP for scratches. If there are scratches on the surface, take a
homogeneous test image (see "Check the PSP with a homogeneous test
image"), replace the PSP, if applicable.
❯ Keep an eye on the image quality with each X-ray image taken, also refer to
"14.1 Poor X-ray image".
❯ Take a homogeneous test image of the PSP (refer to "Check the PSP with a
homogeneous test image"). If scratches or artifacts are visible in the image that
may possibly have an adverse effect on the diagnostics, replace the PSP.
❯ Check X-ray system.
❯ Take an X-ray image with a 2D X-ray test phantom. Check the image for
homogeneity, resolution, contrast and artifacts, refer to "Check X-ray system".
2151100018L40 2012V016