sufficient length of tubing to allow reconnection.
12) Turn the holder fastening knob.
13) Press
the
release
counterclockwise the blue ring (indicated by the
lock open).
14) Remove the oxygenator module.
15) Position a new oxygenator module on the holder.
16) Connect the oxygenator inlet line and the arterial
line to the oxygenator module.
17) Connect and turn on the thermocirculator.
18) If possible, connect the reservoir to the
oxygenator module by turning the blue ring nut
(shown with the padlock closed).
19) Connect the purge line to the turret of the
cardiotomy/reservoir;
cardioplegia line.
20) Connect the gas line, the arterial sampling line
and all the monitoring lines.
21) Remove the clamp from the venous line.
22) Open the recirculation line (filter purge or
arteriovenous loop), fill the oxygenator module
by means of the arterial pump, eliminate the
air in the module by opening the purge line in
sequence, and recirculate at a maximum flow
rate of 1000 ml/min.
23) Close the recirculation line remove the clamp
from the arterial line and start the new bypass.
24) Close the purge line and reconnect the
cardioplegia circuit if necessary.
25) Restore the vacuum in the cardiotomy/venous
reservoir if using the vacuum-assisted drainage
technique.
MEDICAL DEVICES FOR USE WITH TRILLY
PAEDIATRIC AF
There are no contraindications for use of any
heating/cooling system (thermocirculator).
At present EUROSETS is not aware of any
contraindication to use of the device with occlusive
pumps or centrifugal pumps. The use of other types
of pumps must be agreed with EUROSETS.
The circuit connections must be made using tubes
of a diameter compatible with the dimensions of the
ports on the device (3/8", 1/4").
Eurosets recommended to always use the
convenience kit, included for each oxygenator,
consisting of 3 x 3/8"-1/4" reducers ports,
1 x cardioplegia adapter and 1 x venous return
extension.
Use Sechrist or a system with similar technical
features as air/oxygen mixer.
button
and
rotate
reconnect
the
Eurosets recommends the use of YSI temperature
probes.
When using the vacuum-assisted drainage
technique, it is recommended to use the specific
Eurosets connection lines between the cardiotomy/
venous reservoir and the vacuum regulator.
LIMITED WARRANTY
These warranty conditions are in addition to any
statutory rights of the purchaser pursuant to
applicable law.
EUROSETS warrants that all reasonable care has
been taken in the manufacture of this medical
device as required by the nature of the device and
the use for which the device is intended.
EUROSETS warrants that the medical device
is capable of functioning as indicated in these
instructions for use when used in accordance with
them by a qualified user and before any expiry date
indicated on the packaging.
Nonetheless, EUROSETS cannot guarantee that
incorrect diagnosis or therapy and/or the particular
physical and biological characteristics of an
individual patient do not affect the performance
and effectiveness of the device with damaging
consequences for the patient, even though the
specified instructions for use have been respected.
EUROSETS, whilst emphasizing the need to adhere
strictly to the instructions for use and to adopt all
the necessary precautions for proper use of the
device, cannot assume any responsibility for any
loss, damage, expense, incidents or consequences
arising directly or indirectly from improper use of
this device.
EUROSETS undertakes to replace the medical
device in the event that it is defective at the time of
placing on the market or, if shipped by EUROSETS,
at the time of delivery to the end user, unless such
defect is attributable to the purchaser.
The above replaces all other warranties on sale
and/or functionality.
No EUROSETS representative, dealer, distributor or
intermediary or any other industrial or commercial
organisation is authorised to make any claim or
provide additional warranties other than those
expressly stated herein.
EUROSETS declines all responsibility for any
modifications to the Warranty Conditions and the
information/instructions for use expressly stated
herein.
The purchaser undertakes to comply with the terms