Power Cable
Specifications
Overload protection
Electrical
Requirements
Fuse
Power Supply DC
Output Voltage
Electrical Leakage
Environmental
Specifications
Integrated Continuous Battery Charging System
The specifications required by the MEDRAD Spectris Solaris for use with The
Integrated Continuous Battery System relative to the power cable (plug,
receptacle, and cord) are:
•
Operating Temperature: 60° C minimum
•
Receptacle Type: IEC-60320 C13
•
Normal Cord Voltage: 300 VAC minimum
•
Wire Gauge: 1.00 mm2 minimum
•
Cord Type: IEC 60245-1, Annex A, Designation 53, or IEC
60227-1, Annex A, Designation 53 Certified
•
Cord Length: 3 m maximum
The power cable must meet applicable plug, cord, and receptacle
specifications including type, voltage, current, and safety approval markings
for the country in which the power cable is being used.
The Integrated Continuous Battery Charging System provides output overload
and short-circuit protection.
This manual applies to the Integrated Continuous Battery Charging System
for use with MEDRAD Spectris Solaris and MEDRAD Spectris Solaris EP MR
Injection Systems. Read all of the information contained in this section.
Understanding the information will assist you in operating the device in a safe
manner.
The Integrated Continuous Battery Charging System is designed to comply
IEC 60601-1 (2nd and 3rd Edition Amendment 1) and IEC 60601-1-2 (2nd
and 3rd Edition) standards, including national differences.
100-240 VAC
50/60 Hz
100 VA
Fuse 2.5A, 250V, 5X20MM, IEC TYPE F, HIGH
Nominal 17.0 VDC
Unit < 300 microamperes @ 264 VAC 60 Hz
Patient < 10 microamperes
Non-Operating: (Transportation and Storage)
Temperature:
-25 °C to 70 °C (-13 °F to +158 °F)
Humidity:
5 % to 100 % R.H.
Air Pressure:
48 kPa to 110 kPa (6.96 psi-16 psi)
16