Configurations
Status Indicator
Battery Status
Integrated Continuous Battery Charging System
WARNING: Contains Ferromagnetic material. Failure to properly
install the Integrated Continuous Battery Charger system Power
Supply may result in equipment damage or injury. Attach the
power supply to the floor or wall using the provided adhesive
strips at least 6 ft. (1.8 m) from the scanner face.
The Integrated Continuous Battery Charging System is available in two
configurations:
•
Scan Room Configuration: places the power supply in the Scan Room
with the cables routing directly to the Scan Room Unit.
•
Control Room Configuration: places the power supply outside the Scan
Room with the cables routing through the penetration panel into the Scan
Room.
Scan Room
Configuration
NOTE:
Follow all institutional, local, or national safety regulations
related to routing cabling on the floor.
An indicator on the Control Room Unit displays the charging system's
operation status. When this indicator is illuminated yellow, this shows that the
Integrated Continuous Battery Charging system is present and functioning. If
the indicator is not illuminated, verify the power supply is plugged in and the
cables are connected.
WARNING: Explosion Hazard. Serious injury or death may result
from improper use of the Integrated Continuous Battery
Charging System. The Integrated Continuous Battery Charging
System is intended for use in a well ventilated area.
When ARM is pressed, check the status of the battery in the lower left corner
Status Indicator
12
Control Room
Configuration
Penetration
Panel