Device Description
All manuals and user guides at all-guides.com
F2 Lg Washer Port 1st Dilator
F2 Lg Washer Port 2nd Dilator
F2 Sm Washer Port
F2 Lg Washer Port
9F Suction Tube
9F Suction Tube Retractor
11F Suction Tube
11F Suction Tube Retractor
•
Push the brush through the entire length of the lumen using a twisting motion to remove debris
at least five times.
•
Use a 50ml syringe filled to capacity with enzymatic cleaning solution to flush the lumen of
each device.
All Instruments (cont.):
•
Rinse the instrument with warm, 85°F – 104°F (30°C – 40°C), tap water for a minimum of one
minute and until visual evidence of debris, soil, and cleaning solution are gone.
•
Pay particular attention to flush the lumen of the dilators, ports, and suction tubes with warm, 85°F
– 104°F (30°C – 40°C), tap water.
•
Ultrasonically clean the device components for 10 minutes in neutral pH detergent, prepared in
accordance with the manufacturer's instructions.
•
Rinse the device components with warm, 85°F – 104°F (30°C – 40°C), tap water for a minimum of
one minute and until visual evidence of debris, soil, and cleaning solution are gone.
Dry the device components immediately after final rinse with a clean towel or compressed air until
visibly dry.
Automated Cleaning
•
Perform Manual Cleaning prior to Automated Cleaning for SPOTLIGHT Access Ports, Dilators,
and Suction Tubes.
•
Use a soft bristle brush (plastic bristles, like nylon) to thoroughly remove all traces of blood and debris.
•
Load the device components so that the lumens of the ports, dilators, and suction tubes can drain.
•
Clean, using the "INSTRUMENTS" cycle in a validated washer disinfector and a pH neutral
cleaning agent intended for use in automated cleaning using the minimum cycle parameters below:
Phase
Pre-Wash
Enzyme Wash
Wash
Rinse
Dry
Product Code
2757-10-078
2757-10-079
2757-10-074
2757-10-080
2929-08-009
2929-08-109
2929-08-011
2929-08-111
— OR —
Recirculation Time
2:00
1:00
2:00
0:15
15:00
8 of 152
Lumen Size
7.3 mm
13 mm
14.6 mm
16.5 mm
9F or 3 mm
9F or 3 mm
11F or 4 mm
11F or 4 mm
Water Temp
Cold Tap Water
Hot Tap Water
65.5°C
Hot Tap Water
66°C
Brush Diameter
8 mm
14 mm
15 mm
Toothbrush
3 mm
3 mm
4 mm
4 mm
Detergent Type
N/A
Enzymatic Cleaner
Neutral pH Detergent
N/A
N/A