P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need
to perform preventive maintenance checks more often based on your level of product usage.
Remove product from service before performing preventive maintenance.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items for the standard models:
Zipper and covers (top and bottom) are free of tears, cuts, holes, or other openings
Internal components for signs of staining from fluid ingress or contamination by fully unzipping the covers
Labels for legibility, proper adherence, and integrity
Handles are free of rips or cracks and stitching is intact
Foam has not degraded or come apart
Fire barrier for rips, cracks, or other visible signs of damage
Inspect the following items for the behavioral health models:
Covers (top and bottom) are free of tears, cuts, holes, or other openings
Labels for legibility, proper adherence, and integrity
Product serial number:
Completed by:
Date:
C C h h e e c c k k i i n n g g t t h h e e c c o o m m p p r r e e s s s s i i o o n n s s e e t t o o f f t t h h e e f f o o a a m m
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
•
35" + Straight Edge
•
Tape Measure
P P r r o o c c e e d d u u r r e e : :
1. Raise the bed height to the full up position.
2. Lower the fowler and gatch sections to the full down position.
N N o o t t e e - - Before taking the compression set measurement, make sure that there has not been a patient on the support
surface for a minimum of one hour.
3. Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient right
corner.
4. Fold the top of the cover to the patient's right side.
5. If the support surface has the optional fire barrier, starting at the foot end, roll up the fire barrier on the foam crib
assembly.
N N o o t t e e - - Work from side to side a little at a time to get the fire barrier past the seat section of the foam crib.
6. Place a straight edge (at least 35 in. (88.9 cm) long) across the foam crib from left to right at the sacral region (patient
seat section) of the support surface (Figure 1).
7. Using a tape measure, measure the maximum depth (lowest point) from the bottom of the straight edge to the top of the
foam (Figure 2).
2850-109-005 Rev A.0
11
EN