2.7.2.1.
Special Tare Marking for Multi-Interval Weighing Applications
Because of the way the IND570 manages the tare function in multi-interval applications (refer to
section 2.7.2.3 in Operation), it is required that the final weighing system has a label displayed
that indicates the maximum tare value within the capacity of interval # 1: T = nnn, where nnn is the
capacity of interval 1.
The requirements for this marking are as follows:
2.7.2.2.
Label Requirements for MID Weighing Applications
To satisfy the labelling requirements for MID R51 (Automatic Catchweigher) and MID
2.8.

Sealing the Enclosure

When the IND570 terminal is used in a metrologically "approved" application, it must be protected
from tampering by use of seals. An optional sealing kit is available from METTLER TOLEDO. The kit
(part number 30130836) contains all the required hardware. The method used for sealing will vary
depending upon local requirements. The IND570 supports external sealing.
2-72
METTLER TOLEDO IND570 Terminal Installation Manual
Capacity Label
in Place
Required only for OIML approved terminals that are programmed for multi-interval (not multiple
range).
The descriptive markings shall be indelible and of a size, shape and clarity allowing easy
reading.
It shall be located in a clearly visible place on a sticker fixed permanently to the instrument.
In case the sticker is not destroyed when removed, a means of securing shall be provided, e.g.
a control mark that can be applied.
R61(Automatic Gravimetric Filling) applications, an MID Labelset specific to IND570 can be
purchased and applied according to local regulations. Please refer to the User's Guide Chapter
6, Parts and Accessories, for ordering information.
Figure 2-125: Capacity Label Installed
30205319
|
11
|
07/2019