2.4.6.3.2.
Passive Input Wiring
The inputs can be selected as either active or passive via a switch on the board. When "active" is
selected, an external voltage source is required.
2-56
METTLER TOLEDO IND570 Terminal Installation Manual
Notes for Passive Input Discrete I/O wiring:
Voltage: 5-30VDC, 10 mA max. current. Voltage polarity may be reversed
The two IN COM terminals are connected internally on the DI/O pcb so either IN COM can
be utilized during installation. Two IN COMs are provided as a convenience for installations
where external switches/contacts are not located together and multiple commons require
wire termination.
Do not bundle input wiring with power wiring or any other high energy cables
Wire sizes: 22 AWG min./14 AWG max.
Figure 2-102: Passive Input Wiring
30205319
|
11
|
07/2019