3.2 Selecting the location
The correct location is crucial for the accuracy of the weighing results.
1 Select a stable, vibration-free and, if possible, a horizontal location for
the weighing platform.
The ground must be able to safely bear the weight of the fully loaded
weighing platform.
2 Observe the following environmental conditions:
No direct sunlight
No strong drafts
No excessive temperature fluctuations
3.3 Levelling
Levelling of weighing platforms
Only weighing platforms that have been leveled precisely horizontally
provide accurate weighing results. Weights and Measures approved
weighing platforms have a level bubble to simplify levelling.
Turn the adjustable feet of the weighing platform until the level bubble's
air bubble is inside the inner circle.
3.4 Weighing platform connection
Analog weighing platforms
Call the METTLER TOLEDO service technician to connect an analog weighing platform to the ICS4__a /
ICS68_a weighing terminal.
Weighing platforms with digital scale interface
Connect the weighing platform connector to the IDNet or SICSpro scale interface of the weighing terminal.
3.5 Power supply connection
Plug the power plug into the power socket.
After it has been connected, the device runs a self-test. The device is ready to operate when zero appears on
the display.
ICS4__ / ICS68_
CAUTION
Risk of electric shock!
1 Before connecting the power supply, check whether the voltage value printed on the
label corresponds to your local system voltage.
2 Do not, under any circumstances, connect the device if the voltage value on the label
deviates from the local system voltage.
3 Make sure the weighing platform has reached room temperature before switching on the
power supply.
Installation
7