Diagnostics And Maintenance; Common Errors; Cleaning The Terminal - Mettler Toledo IND231 Guía Rápida

Ocultar thumbs Ver también para IND231:
Tabla de contenido

6. Diagnostics and Maintenance

Common Errors

Overcapacity
Undercapacity
- - no - -
Scale in Motion
Failure to Zero
Error 4
Data Too Long
Terminal turns
Terminal is set to auto power-off
itself off

Cleaning the Terminal

Use a soft, clean cloth and mild glass cleaner. Do not spray cleaner directly onto terminal.
Do not use industrial solvents such as acetone.
14
METTLER TOLEDO IND231/IND236 Quick Guide
Terminal cannot execute commands because the weight on the scale is over the
calibrated capacity. The weight display shows blanked condition
Terminal cannot execute commands because the weight is under the current captured
zero. The weight display will show an under zero condition
Incorrect or invalid key-press
If motion is detected when a command is received, the IND231/IND236 will wait for a no-
motion condition. If a stable (no motion) weight condition is reached, the command is
executed. If a no-motion state cannot be reached, the command is aborted and a "Scale In
Motion" icon appears at lower left on the display
Gross weight outside the programmed zero range
In Counting application: Sample size is too small
Data is longer than the screen can display
or
battery voltage is too low
30100793
|
02
|
01/2016
Tabla de contenido
loading

Este manual también es adecuado para:

Ind236

Tabla de contenido