Both of these steps must be correctly completed to ensure the correct function of the PowerMount
modules.
2.4.5.8.1.
Cable and Gland Preparation, Harsh Enclosure
Ground and shield terminations are a critical part of the PowerMount system's immunity to noise
and electrical surges. Prepare and install the terminal end of the PowerMount cable as follows:
2-36
METTLER TOLEDO IND570 Terminal Installation Manual
Internal connection and grounding of the cable.
Figure 2-66: IND570 Harsh Enclosure with Cable Gland for PowerMount Connection
Figure 2-67: Cable Preparation Measurements
Table 2-8: Cable Preparation Lengths – Harsh
Letter
A
Length of inner braided shield
B
Length of red, white and blue wires
C
Length of black wire and drain wire
D
Amount of insulation to strip from wires
Description
Length
32 mm (1.25")
254 mm (10")
32 mm (1.25")
5 mm (0.2")
30205319
|