2.4.5.1.2.
DC Powered Models
24 VDC power is only available on panel mount models. The DC terminal requires 24 VDC, -15%,
+20% (at 1.25A maximum).
No power cables are included with the 24 VDC powered IND570 terminals. The 24 VDC power and
ground must be brought directly to the power connection of the main board and terminated there.
Figure 2-27 and show the three position terminal block provided for the DC power connection. A
wire is not terminated to the center position.
2.4.5.2.
Analog Load Cell Connections
TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE IND570 TERMINAL AND WAIT AT
LEAST 30 SECONDS BEFORE CONNECTING OR DISCONNECTING ANY HARNESS.
When using an analog load cell version of the IND570, load cell connections are made to the
connector located on the main board as shown in Figure 2-25.
30205319
|
11
|
07/2019
Figure 2-30: AC Power Connections on Harsh Models
NOTICE
METTLER TOLEDO IND570 Terminal Installation Manual
AC Ground
Connection
AC Power
Connection
AC Power
Line Label
2-19