Guidelines and Manufacturer's Statement - Electromag-
netic Compatibility
14.1

Impact of Mobile and Portable HF Communication Devices

The emission of high frequency energy by mobile communication devices may
impact the function of the electrical medical device. Operating such devices (e.g.,
cell phones, GSM phones) in the proximity of the electrical medical device is pro-
hibited.
14.2

Electrical Connections

Do not touch electrical connections identified with this warning label. Do not es-
tablish a connection between these plugs and sockets without first implement-
ing precautionary ESD (electrostatic discharge) measures.
The following are ESD precautionary measures:
• Apply potential equalization (PE), if available on your equipment, to all devices
to be connected.
• Use only the listed equipment and accessories.
Hospital employees should be informed about and trained in ESD precautionary
measures.
14.3

Accessories

An autoclavable remote control (cable length < 4.5 m / 177.2 in), can be connect-
ed to the FloSteady Arthroscopy Pump.

Guidelines and Manufacturer's Statement - Electromagnetic Compatibility

ESD (Electrostatic Discharge) precautionary
measures
Remote control
EN
51