13 - Preparing the ABPM 7100
Technical Data and environmental conditions
Measurement pressure range:
Systolic
60 to 290 mmHg
Diastolic
30 to 195 mmHg
Accuracy:
+/- 3 mmHg in display range
Static pressure range:
0 to 300 mmHg
Pulse range:
30 to 240 beats per minute
Procedure:
oscillometric
Measurement intervals:
0, 1, 2, 3, 4, 5, 6, 10, 12, 15, 20 or 30 measurements per hour
Measurement logs:
4 adjustable interval groups
Memory capacity:
300 Measurements (with CBP or PWA: 260 measurements)
Battery capacity:
> 300 Measurements
Operating temperatures:
+10°C to +40°C
Operating humidity:
15% to 90% rel. humidity
Storage environment:
-20°C to +50°C and 15% to 95% rel. humidity
Ambient pressure
700-1060 hPa
Dimensions:
121 x 80 x 33 mm
Weight:
approx. 220 g excluding batteries
2 Ni-MH batteries with 1,2 V each and min. 1500 mAh (AA, Mignon) or 2
Power supply:
Alkali 1,5 V batteries (AA, Mignon, LR6)
Interfaces:
USB-Interface cable
Expected operational device life:
5 Years
Expected operational cuff life:
6 Months
Environmental conditions:
Attention!
•
Extremes of temperature, humidity or air pressure can affect measurement accuracy. Please observe the
operating conditions.
•
Extreme temperatures, humidity or altitude can affect the performance of the blood pressure monitor. Do not
store the device near a fireplace or heating unit and do not expose it to intense sunlight. Do not place the
device near a nebuliser or steam generator, as the condensation may damage it.
•
The blood pressure monitor takes approx. 25 minutes to go from the minimum storage temperature of
-20°C to the operating temperature of +10°C in an ambient temperature of 20°C.
•
The blood pressure monitor takes approx. 25 minutes to go from the maximum storage temperature of
+50°C to the operating temperature of +40°C in an ambient temperature of +20°C.
Accessories
Accessories
Name
REUSE-09-ABPM
CUFF, WA, REUSE, CHILD, ABPM
CUFF, WA, REUSE, SMALLADULT,
REUSE-10-ABPM
ABPM
REUSE-11-ABPM
CUFF, WA, REUSE, ADULT, ABPM
CUFF, WA, REUSE, ADULTPLUS,
REUSE-11L-ABPM
ABPM
CUFF, WA, REUSE, LARGEADULT,
REUSE-12-ABPM
ABPM
REUSE-091012-
CUFF, WA, REUSE, CSL, ABPM
ABPM
MANSCHETTE, WA, REUSE, ALL,
REUSE-ALL-ABPM
ABPM
7100-21
ABPM 7100 POUCH PACK
ABPM 7100_CABLE_PC_
7100-24
INTERFACE
ABPM 7100 BATT COVER
7100-10
REPLACEMENT
ABPM-7100CBP-
ABPM 7100 UPGRADE KIT CBP
UPGRA
IEM
ABPM-7100PWA-
ABPM 7100 UPGRADE KIT PWA
UPGRA
IEM
CBP-TO-PWA-
ABPM 7100 UPGRADE KIT CBP to
UPGRA
PWA IEM
Product description - 14
Product description
Reusable cuff for 24-hour blood pressure
measurement in children
(Arm circumference: 14-20 cm [5.5-7.9 inches])
Reusable adult cuff size S for
24-hour blood pressure measurement
(Arm circumference: 20-24 cm [7.9-9.5 inches])
Reusable adult cuff size M for
24-hour blood pressure measurement
(Arm circumference: 24-32 cm [9.5-12.6
inches])
Reusable adult cuff size L for
24-hour blood pressure measurement
(Arm circumference 32-38 cm [12.6-15.0
inches])
Reusable adult cuff size XL for
24-hour blood pressure measurement
(Arm circumference: 38-55 cm [15.0-21.7
inches])
Box of reusable cuffs for 24-hour blood
pressure measurement CHILD, SMALLADULT
and ADULTPLUS
Box of reusable cuffs in all 5 sizes for 24-hour
blood pressure measurement
Carry case for the ABPM 7100 for 24-hour
blood pressure measurement
USB connector cable for computer
Replacement battery compartment cover
ABPM 7100 upgrade to determine central
blood pressure values (monitor serial number
required)
ABPM 7100 upgrade for pulse wave analysis
(monitor serial number required)
ABPM 7100 upgrade with CBP upgrade to
PWA pulse wave analysis device (monitor
serial number required)
EN