Patient Information - operation of the ABPM 7100
Measurement process
During the first measurement, the cuff is inflated in increments, to determine the cuff pressure required to measure
the systolic blood pressure value. This maximum required inflation pressure is stored and applied by direct inflation
during the subsequent automatic measurements. The patient should stay calm during the measurement process,
until the measurement is completed. Allow your arm to hang loose, or place your lower arm loosely on the table or
on a support whilst sitting. Avoid any movement! In the event of a failed measurement a new measurement is
performed automatically according to the measurement process described above.
Cancelling a measurement
A measurement will be cancelled by pressing any buttons during the measurement process causing the cuff to be
quickly deflated automatically. The LCD display will then show "-StoP-" and the ABPM 7100 will beep 5 times. This
cancellation will be stored in the measurement value table under Cancel.
Acoustic signals
The acoustic signals produced by the device consist of single or multiple tones. The following tones are produced:
Tone
Heard during
1 tone
Power on and off
Start and end of measurement (except during night-time interval)
Removal of interface cable
Measurement error
3 tones
System error
Continuous tone
Serious system error (e.g. cuff pressure outside of measurements is over 15
mmHg for more than 10 seconds)
Troubleshooting
An error code appears on the display of the ABPM 7100 for a few seconds in the event of a measurement or system
error. The help information below explains what to do for each error code:
Error code
Action
Keep your arm still during the measurement. If the error recurs, refit the cuff.
ERR 1
If the error appears continuously, please contact your doctor.
Keep your arm still during the measurement. If the error recurs repeatedly,
check the position of the cuff and ensure that the cuff tube is securely
ERR 2
connected to the ABPM 7100.
Keep your arm still during the measurement. If this error persists, please
ERR 3
contact your doctor.
The batteries of the ABPM 7100 are flat. Please contact your doctor.
ERR 5
Check whether the cuff tube is kinked. If the error persists, please contact
ERR 6
your doctor.
ERR 7
The measurement memory is full. Please contact your doctor.
The measurement was terminated because a button was pressed. Repeat
ERR 8
the measurement. Press the START button.
Please contact your doctor.
ERR 9
Please contact your doctor.
ERR 10