39 - EMC Guidelines and Manufacturer Declaration
Table 2 – Guidelines and Manufacturer Declaration
Electromagnetic immunity – for all ME devices and ME systems
Guidelines and Manufacturer Declaration - Electromagnetic immunity
The ABPM 7100 is intended for operation in an electromagnetic environment as specified below. The customer or
user of the ABPM 7100 should ensure it is used in such an environment.
Immunity tests
Test level
Compliance levels
Electrostatic discharge
± 8 kV
± 8 kV
(SD)
Contact discharge
Contact discharge
according to IEC
61000-4-2
± 15 kV
± 15 kV
Air discharge
Air discharge
Rapid transient
± 1 kV
± 1 kV
electrical
100 kHz repetition
100 kHz repetition
disturbance/bursts
rate
rate
according to IEC
61000-4-4
Surges
± 1 kV
Not applicable
according to IEC
Line-to-line voltage
61000-4-5
± 2 kV
Line-to-earth voltage
Not applicable
Magnetic field in
30 A/m
30 A/m
supply frequency
(50/60 Hz)
according to IEC
61000-4-8
Voltage drops,
0% UT
Not applicable
short interruptions
for 0.5 cycles
and fluctuations in
supply voltage
0% UT
Not applicable
according to IEC
for 1 cycle
61000-4-11
70% UT
Not applicable
for 25/30 cycles
0% UT
Not applicable
for 250/300 cycles
NOTE UT is the AC voltage before the application of the test levels.
Electromagnetic Environment -
Guidelines
Floors should consist of wood or cement
or ceramic tiles. If the floor consists of
synthetic materials, relative humidity must
be at least 30%.
The ABPM 7100 does not have an AC
power supply.
Magnet fields in network frequency should
correspond with the typical values found
in business and hospital environments.
The ABPM 7100 does not have an AC
power supply.
Table 3 - Electromagnetic immunity for casings designed to shield against high-frequency wireless
communication devices
Guidelines and manufacturer declaration – electromagnetic immunity
The ABPM 7100 is intended to be operated in the electromagnetic environment specified below. The customer
or the ABPM 7100 user should ensure that it is used in such an environment.
Emitted interference
Test level
measurement
HF radiated disturbances in
380 - 390 MHz
accordance with IEC 61000-4-3
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
Table 4 – Guidelines and Manufacturer Declaration
Electromagnetic immunity for ME devices or ME systems that are not life-supporting
Guidelines and Manufacturer Declaration - Electromagnetic immunity
The ABPM 7100 is intended for operation in an electromagnetic environment as specified below. The customer or
user of the ABPM 7100 should ensure it is used in such an environment.
Immunity tests
Test level
Radiated disturbance variables
10 V/m
according to
80 MHz to
IEC 61000-4-3
2.7 GHz
Conducted disturbance variables
according to
IEC 61000-4-6
EMC Guidelines and Manufacturer Declaration - 40
Compliance level
380 - 390 MHz
27 V/m; PM 50%; 18 Hz
430 - 470 MHz
28 V/m; (FM ±5 kHz, 1 kHz sine)
PM; 18 Hz
704 - 787 MHz
9 V/m; PM 50%; 217 Hz
800 - 960 MHz
28 V/m; PM 50%; 18 Hz
1700 - 1990 MHz
28 V/m; PM 50%; 217 Hz
2400 - 2570 MHz
28 V/m; PM 50%; 217 Hz
5100 - 5800 MHz
9 V/m; PM 50%; 217 Hz
Compliance level
10 V/m
Not applicable