29 - Care and Maintenance

Care and Maintenance
To ensure the optimal functionality of the ABPM 7100 regular care and maintenance of the unit is required.
Attention
Damage to device
Do not open the casing. Once the device is opened, all warranties will lapse.

Cleaning

Cleaning the ABP Monitor and carrying pouch
1. Read the safety instructions carefully and observe them closely before cleaning.
2. Only use a cotton cloth moistened with lukewarm water and mild detergents to clean the ABPM 7100 and the
pouch.
Attention
Damage to the ABP Monitor and carrying pouch caused by the use of solvents
Do not use strong or solvent-based additives.
Ensure that no liquid enters the device.
If liquid does penetrate the device, switch if off immediately and return it to your Welch Allyn specialist for
inspection.
Cleaning the cuff sleeve, bladder and tubing
1. Read the safety instructions carefully and observe them closely before cleaning.
2. Before washing, carefully remove the bladder and tubing from the cuff sleeve.
3. When cleaning the cuff sleeve, bladder and tubing, use only mild detergents in lukewarm water without fabric
softener.
Attention
Damage to the cuff sleeve during washing
Always close the Velcro strip before washing!
It is possible to wash the cuff sleeve in the washing machine at max. 30°C. Do not spin.
Do not use fabric softeners or other washing aids (e.g. hygiene rinses, textile deodorants). These agents can
leave behind residues and damage the material.
The cuff sleeve is not suitable for drying in a dryer.

Disinfection

Caution
Intolerances caused by the use of disinfectants: Some patients display intolerance (e.g. allergies) to
disinfectants or their components.
Never use disinfectants which leave residues on the product or which are not suitable for contact with the
skin.
Carefully wash the cuff to remove residues.
Attention
Damage to the cuff sleeve, bladder and tubing caused by disinfectants
Do not submerge the cuff sleeve in disinfectants.
Avoid disinfecting the cuff bladder and connected rubber tubing.
The bladder and tubing can be damaged by disinfectants. Wipe down the bladder with lukewarm water and
add a mild detergent, if necessary.
Ensure that no liquid enters the tube opening.
The user (doctor) decides whether and when the ABP Monitor and the cuff sleeve should be disinfected for hygienic
reasons (e.g. after every use).
The following agents are recommended for disinfecting the ABP Monitor and the cuff sleeve:
Terralin Liquid (Manufacturer: Schülke & Mayr)
Promanum N (B. Braun)
For full effectiveness, moisten the ABPM 7100 and cuff sleeve with the disinfectant for at least 5 minutes.
The use of disinfectants not recommended in the instructions for use shall render the user responsible for proof of
safe application.
Note
It is imperative that you observe the manufacturer's information regarding the use of these
products. Allow the agents to dry off completely.
Care and Maintenance - 30
EN