14 Malfunctions
Event
Possible cause
10267
Clock-controlled evacuation. The pressure
change is less than expected; the vacuum
performance is decreasing.
10268
Steam intake. The pressure change is less
than expected; the steam intake
performance is decreasing.
10269
Ventilation. The volume flow during
ventilation is lower than expected.
10270
Pressure release. The pressure change is
less than expected; the pressure release
speed is decreasing.
64
What you can do
Please check the dust filter for contamination
and replace it if necessary.
The intake area of the cooling system under
the device must be free; check for paper or
similar items underneath the device which
could stop the air flow.
Please ensure that the device is sufficiently
ventilated. Heat dissipation must be able to
take place freely, ensure ventilation;
conversion cabinets are not recommended.
Please observe the set-up conditions (e.g.
ambient temperature).
Check the loading of the device for
compliance with the permissible loading
quantities.
Check the pressure release filter in the
chamber for blockage.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Check the loading of the device for
compliance with the loading quantities.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Check the sterile air filter in the media board;
if it is heavily worn or blocked, replace it.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Please check the dust filter for contamination
and replace it if necessary.
The intake area of the cooling system under
the device must be free; check for paper or
similar items underneath the device which
could stop the air flow.
Please ensure that the device is sufficiently
ventilated. Heat dissipation must be able to
take place freely, ensure ventilation;
conversion cabinets are not recommended.
Please observe the set-up conditions (e.g.
ambient temperature).
Check the loading of the device for
compliance with the permissible loading
quantities.
Check the pressure release filter in the
chamber for blockage.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.