Changing the dust filter
The following must be fulfilled or present:
ü
A new and dry dust filter.
1.
Open the service hatch.
2.
Press down the centre of the grip and pull out the dust filter.
3.
Insert the new dust filter until it snaps into place.
The latch nose of the grip must point upwards.
4.
Close the service hatch.
Cleaning
NOTICE
Inappropriately performed cleaning can lead to the scratching of and damage to surfaces and the
development of leaks in sealing surfaces.
This also favours the development of soiling deposits and
Comply with all information regarding cleaning of the part affected.
n
Do not use any hard objects for cleaning such as a metal saucepan cleaner or a steel brush.
n
Sterilization chamber, chamber sealing face, mount, trays
To maintain the value of your device and to avoid persistent contamination and deposits, MELAG recommends weekly
cleaning of the surfaces (e.g. with the MELAG chamber cleaning set).
Cleaning the sterilization chamber, chamber sealing face, mount and trays
The following must be fulfilled or present:
The device has been switched off and the power plug has been unplugged from the socket.
ü
ü
The device has been completely cooled.
Trays or sterilization containers and the associated mount have been removed from the sterilization chamber.
ü
1.
Wet the surfaces to be cleaned completely with the cleaning fluid.
Please note: You should not allow cleaning fluid to enter the piping coming from the sterilization chamber.
2.
Spread the cleaning fluid evenly with a non-fuzzing cloth.
3.
Allow the cleaning fluid to act and evaporate for a sufficient time.
4.
Using a new non-fuzzing cloth, spread ample amounts of demineralised water over the cleaned surfaces.
5.
Wipe the surfaces thoroughly to remove cleaning residues. Repeat this process as necessary after wringing out the
cloth.
Ê
Residues of cleaning fluids can ignite or cause deposits on the instruments.
6.
Allow the cleaned areas to dry completely. This may take a few minutes.
7.
Finally, wipe the cleaned surfaces with a dry, non-fuzzing microfibre cloth.
corrosion
in the
sterilization
}
}
12 Maintenance
chamber.