13 Pause times
13 Pause times
Starting after long pauses
After a long period of non-use, the device will heat up after being switched on.
Decommissioning
When decommissioning the device for a long pause (e.g. due to holiday), proceed as follows:
1.
Empty the double jacket steam generator, see
2.
Switch on the device at the power switch.
3.
Disconnect the power plug from the socket and if necessary, allow the device to cool.
4.
Empty the internal storage tank via the drain hose.
5.
Shut off if present, the water inflow of the water treatment unit.
Emptying
You have the option of emptying the water in the double jacket steam generator easily via the Draining program. In order
to do so, the device is heated once, building up pressure in the double jacket so that the water can be emptied fully from
the double jacket steam generator.
Working in the Service programs menu, select the Draining program and press Start.
1.
2.
Confirm the dialogue window.
Ê
The double jacket steam generator is emptied.
3.
Confirm the message Draining successful.
Ê
The device switches to door mode.
4.
Switch off the device.
56
Emptying
[} page 56].