Event
Possible cause
10165
Double jacket emptying. The maximum
runtime for emptying has expired.
10169
Please contact technical services
10179
Chamber pressure release. The pressure
release valve had to be opened several
times before a pressure drop occurred.
10184
In the program run, the maximum filling level
of the external wastewater container is
reached during manual water disposal.
10185
The external wastewater container is full.
10186
With a manual water supply, the filling level
of the feed water tank is undershot.
10224
There is not enough feed water tank in the
tank.
10226
There is not enough feed water tank in the
tank.
10241
Pressure-controlled evacuation. This is
triggered when the termination gradient is
violated in a gradient-monitored process,
which leads to an abort of the program run
(in the context of pressure-controlled
evacuation/pressure gradient monitoring).
The vacuum performance is insufficient.
What you can do
Please allow the device to cool down.
Please observe the set-up conditions.
Please ensure sufficient ventilation of the
device.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Please contact technical services.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Check the chamber to see if any debris from
loading or packing is clogging the fittings.
Remove and check the coarse filter at the
back bottom of the chamber for blockages.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Please empty the wastewater container
before starting a new program. The
wastewater container can still receive the
wastewater from the currently running
program.
For the Empty program, the wastewater
container must be empty. Please empty the
external wastewater container.
Please fill up the feed water tank.
Please fill the feed water tank up to the
maximum marking of the level increase
before the next program start.
Please fill the feed water tank up to the
maximum marking of the level increase.
Please check the dust filter for contamination
and replace it if necessary.
The intake area of the cooling system under
the device must be free; check for paper or
similar items underneath the device which
could stop the air flow.
Please ensure that the device is sufficiently
ventilated. Heat dissipation must be able to
take place freely, ensure ventilation;
conversion cabinets are not recommended.
Please observe the set-up conditions (e.g.
ambient temperature).
Check the loading of the device for
compliance with the permissible loading
quantities.
Check the pressure release filter in the
chamber for blockage.
14 Malfunctions
61