5 First steps
5 First steps
Setup and installation
PLEASE NOTE
Comply with the specifications of the technical manual during setup and installation. This contains all
building-side requirements.
Record of installation and setup
The record of installation and setup is to be completed by the responsible stockist and a copy sent to MELAG as proof of
the correct setup, installation and initial commissioning. This is a constituent part of any guarantee claim.
Feed water supply
Steam sterilization requires the use of
specifies the guideline values to be observed.
The steam sterilizer requires approx. five litres of feed water for the first filling of the steam sterilizer.
Use of a water treatment unit
A water treatment unit is connected to the drinking water supply. This obviates the need to fill the storage container. The
respective system is selected in accordance with the number of sterilization runs per day and the type of the load. Every
MELAG steam sterilizer can be fitted with a water treatment unit.
PLEASE NOTE
Should you wish to use a water treatment unit from another manufacturer, please consult MELAG.
Use of external storage and wastewater container
Fill the storage container with feed water. The feed water is pumped into the device via the filling pump. Check the water
level in the storage container before every program start. The wastewater container has a capacity of approx. 25 litres
and is sufficient for at least ten sterilizations.
NOTICE
Danger of algae development
To prevent the development of algae, never expose the storage container to sunlight.
n
Switching on the device
The following must be fulfilled or present:
ü
The device is connected to the power supply.
The feed water supply is secure.
ü
ü
The wastewater disposal is connected.
1.
Switch on the device at the power switch.
Ê
The double jacket steam generator is ventilated, whereby the air flow can cause a hissing noise.
2.
The welcome screen is displayed. The display then switches to the Favourites menu.
Ê
The feed water level is checked and pre-heated immediately after activation.
20
distilled
or
demineralized
}
}
water, known as
feed
water. Annex C of
}
EN 13060
}