This may be very hazardous to health. The tube may also be destroyed or damaged.
Depending on the individual needs of the patient, which must be determined by the treating physician,
it may be necessary to immerse the tracheostomy tube in boiling water (212 degrees Fahrenheit/100
EN
degrees Celsius) for at least 5 minutes but not more than 10 minutes.
WARNING
Let the tube/button cool before reinsertion to avoid scalds and burns.
Cleaning Steps
Remove any additional items before cleaning.
First, rinse the tube thoroughly under running water (see picture 7).
Use only lukewarm water to prepare the cleaning solution and observe the instructions for use of the
cleaning agent.
We recommend the use of the cannula-cleaning tub with sieve insert (REF 31200) to facilitate the
cleaning procedure.
To do this, hold the sieve insert by the upper edge to prevent contact with and contamination of the
cleaning solution (see picture 8).
Place only one tracheostomy tube at a time in the sieve insert of the cannula-cleaning tub. If more than
one tube is cleaned at a time, they may be compressed too strongly and damaged by excessive pressure.
After the immersion time has expired (see instructions for use of the cannula cleaning powder), wash the
tracheostomy tube thoroughly several times with lukewarm, clear water (see picture 7). There must be no
residues of the cleaning agent on the tube when it is inserted into the tracheostoma.
If necessary, for example, if persistent secretion residues cannot be removed by the cleaning bath,
additional cleaning with a special cannula-cleaning brush (OPTIBRUSH
, REF 31850 or OPTIBRUSH
®
®
Plus with fibre top, REF 31855) may help. Only use the cleaning brush, if the tube is removed and already
outside the tracheostoma.
Always insert the cannula cleaning brush into the cannula from the cannula tip (see picture 9).
Use the brush as directed and proceed with great care to avoid damage to the soft tube material.
Thoroughly rinse the tube or button under lukewarm running water or sterile saline solution (0.9% NaCl
solution).
Dry the tube thoroughly with a clean and lint-free cloth after wet cleaning.
Never use tracheostomy tubes with impaired functionality or with damages such as sharp edges or
cracks, because this may lead to injuries of the mucus membranes in the trachea. If damages are found,
discard the tube.
2. Chemical Disinfection Instructions
The Fahl
tracheostomy tube can be disinfected by cold disinfection with special chemical disinfectants.
®
Disinfection should always be done if ordered by a doctor/physician due to specific health concerns
caused by disease, infection, or your specific situation.
Disinfecting is generally indicated to prevent cross-infection and in inpatient situations (e.g. hospitals,
nursing homes, and/or other health care facilities) to limit infections.
CAUTION!
Always clean according to the cleaning procedure described above before proceeding to
disinfection (if applicable).
Disinfectants that release chlorine or that contain strong alkalis or phenol derivatives must under
no circumstances be used. This may seriously damage or even destroy the tube or button.
For this purpose, OPTICIT
tube disinfectant (REF 31180) should be used in accordance with the
®
manufacturer's instructions.
As alternative, we recommend a disinfectant based on glutaric aldehyde as active ingredient. Always
observe the area of application and spectrum of activity specified for the disinfectant by the manufacturer.
Follow the instructions for the disinfectant.
Dry the tube thoroughly with a clean and lint-free cloth after wet cleaning.
X. STORAGE
Currently not used and cleaned tubes should be stored in a clean plastic container, dry environment away
from dust, sunlight and heat.
We highly recommend keeping at least two replacement tubes on hand to ensure continuous supply.
XI. SERVICE LIFE
These tracheostomy tubes are products intended for single-patient use.
When used and looked after correctly, in particular observing the instructions for use, cleaning,
disinfection and maintenance provided in the instructions for use, these tracheostomy tubes have an
expected service life of three to six months.
The maximum period of use should not exceed 6 months.