open and locked.
12. Ultrasound probe – not
secure in cable clamp
or comes into contact
with chamber wall
13. The device is failing
cycles.
14. The device takes too
long to warm up
between cycles
15. I cannot locate the
serial number for
trophon EPR
®
trophon
EPR User Manual – N00020
screen
At all other times, power the device OFF using the power
switch and back ON (Figure 4, item 10)
The device should then automatically unlock the chamber
door.
If the above are not successful, power OFF the device,
and follow section B5.3 for the manual door lock
override. Close the Chamber door before powering the
trophon EPR ON again.
Refer to trophon Validated Probes list to confirm probe
compatibility with trophon EPR
Contact your customer service representative.
Record any error codes and colour of the indicator and
contact your customer service representative.
Ensure that the probe is removed immediately after the
end of each cycle.
The trophon EPR serial number can be located on the
back label of your device and also on the side label
above the power socket.