The following are image quality settings that a user can configure. For a more
complete description of these settings refer to Appendix B.
Sharpening
Brightness
Red Saturation
Blue Saturation
Peak / Mean
To adjust an image quality setting:
72
From 0 to 10
From 0 to 255
From 0 to 1024
From 0 to 1024
Peak
Mean
Balanced
Press the Menu button on the front panel of the VPI or select the VPI Options
menu in the Tools menu
On the menu, select Image Quality.
Press the Up or Down arrow button to highlight the desired image setting.
Press the Right arrow button or the
Press the Up or Down arrow button to adjust the setting. The setting takes
effect immediately and is visible in the image.
4-0002428 Rev. J
Larger values increase image
sharpening
Larger values increase image
brightness
Larger values increase red saturation—
red objects look more deeply red
Larger values increase blue
saturation—blue objects look more
deeply blue
Select Peak to set the brightness for
viewing small, foreground objects that
are closer to the front of the
laparoscope.
Select Mean to set the brightness by
weighting the adjustment more by the
mean brightness of the scene and less
by the brightest points.
Select Balanced to set the brightness
by balancing the Peak and Mean
settings.
.
button to select.