CAUTION:
Strictly adhere to the instructions provided by the cleaning agent/disinfectant
manufacturer to determine concentration and exposure time.
The pre-cleaning steps described above should always be performed prior to
cleaning laparoscopes and light guide cables.
After preparation and following the General Practice guidelines, clean
laparoscopes and light guide cables according to the following detailed manual
cleaning or detailed automated cleaning instructions.
WARNING:
Disinfecting the product is NOT a substitute for sterilization as part of the
reprocessing process. The product must be sterilized prior to use on a
patient.
Detailed manual cleaning instructions:
84
Use freshly prepared solutions only.
Use sterile or low-bacteria (up to 10 bacteria/ml) and low-endotoxin (maximum
0.25 EU/ml) water (e.g., purified or highly purified water).
Use filtered air for drying.
Corrosion on the product can occur from using chloride-containing water.
Therefore, fully demineralized water should be used for the final rinse
procedure.
The cleaning agents used must be dispensed according to the prescription of
the manufacturer.
Prepare a fresh solution of enzymatic detergent
Wipe the entire surface of the device using a soft clean cloth dipped in the
detergent solution.
Immerse the device in the detergent solution, ensuring the solution contacts all
inner and outer surfaces.
Soak the device in the solution according to the manufacturer's
recommendations.
3
With the device immersed in the solution, thoroughly brush the exterior with a
soft-bristled brush, focusing on any mated or rough surfaces.
Actuate and brush any movable parts in all extreme positions.
Rinse each device with water
Once all detergent residue is removed, continue to rinse for 30 seconds.
Drain excess water from the device and dry it using a clean cloth or
pressurized air.
4-0002428 Rev. J
2
with lukewarm water
1
until all detergent residue is removed.
1
.