WARNING:
Follow the instructions in Chapter 7 to clean and sterilize the PINPOINT
components and accessories.
Keep the light guide cable connectors clean at all times. Contaminants on the
light guide cable connectors may cause overheating.
WARNING:
For storage, transport and processing, ensure that the laparoscope is not
subjected to mechanical strain to prevent damage to the sensitive lens
system.
The laparoscopes are delivered non-sterile as reusable products.
In general, users are responsible for validation of their re-processing
processes.
Ensure that the processing, material and personnel are suitable for achieving
the results necessary.
Observe all local regulations for operator safety, protection and training
in performance of manual cleaning and drying processes.
Clean and sterilize the laparoscope prior to initial use as well as each
subsequent use of the laparoscope.
Observe appropriate protective measures to prevent contaminating the
environment.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
9