The correct output formats are as follows:
▪
Stryker® SDC3 Information Management System: 3G-SDI
▪
Sony 1000MD recorder: HD-SDI
▪
Sony 3300MT recorder: 3G-SDI
▪
TEAC UR-4MD recorder: HD-SDI or 3G-SDI
▪
MediCapture USB300 recorder: HD-SDI or 3G-SDI
WARNING:
If the video output is routed through a video recorder, operators must always
be aware of whether they are viewing a live or recorded image. Wave your
hand in front of the camera to determine whether or not you are viewing a
live image.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual