To start and stop video recording:
Press and hold the Display Mode button
In the Tools menu, select the Record icon (see Figure 35).
Press the the Display Mode button
A recording icon
recording (see Figure 36).
A recording-stopped
when recording is stopped.
To pause video recording, press the VIDEO/PAUSE button on the front panel
of the recorder. A recording-paused icon
resume recording, press the VIDEO/PAUSE button again.
To enable playback, press the CLOSE button on the front panel of the
recorder.
If the record icon
may not be selected in the VPI Service menu. See Video Recording and
Screenshot Errors on page 105.
Note:
For more information about using the TEAC UR-4MD recorder, see the device
operating instructions.
Note:
This button function is disabled by default and must be enabled in the VPI Options
menu after the PINPOINT System is installed.
Press the rear focus
recording icon displays on the monitor.
When using a Stryker SDC3 system or Sony recorder, press the rear focus
button again to pause recording. When using a TEAC recorder, press the rear
focus button again to stop recording.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
to start and pause recording.
displays in the top right corner of the screen during
icon displays in the top right corner of the screen
displays briefly and then disappears, the TEAC recorder
button to start recording video (see Figure 38). A
to activate the Tools menu.
displays on the screen. To
63