Move the PINPOINT system into an appropriate position for the clinical
imaging procedure.
Plug the VPI power cord into an AC wall outlet or isolation transformer.
The power indicator on the front of the VPI above the VPI power button should
illuminate amber.
▪
▪
Connect the PINPOINT camera to VPI by plugging the camera and light guide
cables into the appropriate ports on the VPI front panel (see Chapter 2 and
Figure 13).
CAUTION:
For the PINPOINT camera to initialize correctly, it must be connected to the VPI,
prior to powering on the system.
Press the power button
The power indicator turns green and the system enters Standby mode. The
Standby mode icon
(see Figure 14).
Note:
In Standby mode, video display is enabled and ventilation fans are on. Illumination
is not active and the video image on the display may be dark.
As the VPI powers on, the PINPOINT System emits dim flashes of colored light.
PINPOINT Endoscopic Fluorescence Imaging System Operator's Manual
If the power indicator does not illuminate, check to confirm that the main
power switch (on the rear panel of the VPI) is turned to the 'on' position.
If, after confirming that the main power switch is in the 'on' position, the
indicator still does not illuminate, please consult Appendix A:
Troubleshooting.
on the front panel of the VPI.
displays in the top-right corner of the video image
41