Storz UNIDRIVE S III Serie Manual Del Usuario página 163

UNIDRIVE
S III
®
D Connect the footswitch.
✓ The error message E12 (no motor) is displayed.
D Connect a handpiece to the motor socket M1.
✓ The handpiece is recognized and the corresponding information displayed. e.g. P0, P1, etc.
D Press the right footswitch pedal.
✓ The motor starts.
D Release the footswitch pedal.
✓ The motor stops immediately.
D Change the direction of rotation by pressing the mid button of the footswitch.
✓ The displayed direction of rotation changes.
For devices with motor socket M2:
D Connect a handpiece to the motor socket M2.
D Select the handpiece M2 by pressing the M button.
✓ The handpiece is recognized and the corresponding information displayed.
D Press the right footswitch pedal.
✓ The motor starts.
D Release the footswitch pedal.
✓ The motor stops immediately.
D Change the direction of rotation by pressing the mid button of the footswitch.
✓ The displayed direction of rotation changes.
For devices providing a pump:
D Press the left footswitch pedal.
✓ The pump starts running.
D Change the flow of the pump pushing the +/- buttons on the display.
✓ The pump has to run regularly at all settings.
D Open the tensioning lever of the pump.
✓ The error message E16 (irrigation pump not closed) is displayed.
D Close the tensioning lever of the pump again.
✓ The error message disappears.
4-5
05.13 / V1.0

Capítulos

Solución de problemas

loading