101
Anhang
Guidance and manufacturer's declaration – electromagnetic immunity
The UNIDRIVE
®
S III ENT model 40 7016 20-1 is intended for use in the electromagnetic environment specified below.
The customer or user of the UNIDRIVE
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
acc. to IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
acc. to IEC 61000-4-4
Surge
± 1 kV differential mode
acc. to IEC 61000-4-5
± 2 kV common mode
Voltage dips, short interrup-
tions and voltage variations
<5% U
*
T
on power supply input lines
(>95% dip in U
acc. to IEC 61000-4-11
for 0.5 cycle
40% U
T
(60% dip in U
for 5 cycles
70% U
T
(30% dip in U
for 25 cycles
<5% U
T
(>95% dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
acc. to IEC 61000-4-8
Appendix
Table 202
®
S III ENT model 40 7016 20-1 should ensure that it is used in such an environment.
Compliance level
6 kV contact
8 kV air
2 kV/1 kV
2 kV/1 kV
0 V, 10 msec
)
T
92 V, 100 msec
)
T
168 V, 500 msec
)
T
0 V, 5 sec
)
T
3 A/m
Anexo
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment.
If the user of the equipment or system requires continued
operation during power mains interruptions, it is recom-
mended that the equipment or system be powered from
an uninterruptible power supply.
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.