Preventative Maintenance

4.8.1 NOTE: Perform annual preventative maintenance procedures to ensure all Cart
components are functioning properly, including but not limited to: (1) electrical power cords for
fraying, damage, and proper grounding; (2) AC power switch; and (3) caster damage.
4.9

Safety Information

4.9.1 PRECAUTION: To avoid electrical hazards, all parts of the ORLab, including the
computer, monitor, Stimulus Generator Unit, and Lexmark laser printer must be connected to the
Isolation Transformer and never directly to a mains outlet.
4.9.2 PRECAUTION: To avoid risk of electric shock, the Isolation Transformer must only be
connected to a supply mains with protective earth.
4.9.3 PRECAUTION: To avoid electrical hazards and prevent overloading the transformer, do
not connect any non-AtriCure approved equipment to the supplied Mains Isolation Transformer.
4.9.4 PRECAUTION: Use AtriCure
flammable gasses. To avoid risk of explosion, the Basic Cart should only be used in a ventilated
area as gasses may be released during charging of backup battery, and should not be used in
rooms with flammable anaesthesia.
4.9.5 PRECAUTION: AtriCure
flammable anaesthetic mixture with air or with oxygen or nitrous.
4.9.6 PRECAUTION: Secure tank with chain and ONLY use N
diameter that weigh ≤ 29.5 kg (65 pounds). Due to the weight of the N
while lifting and securing the tanks.
4.9.7 PRECAUTION: If the cryoICE Box is connected to a grounded hospital AC power
receptacle, it must be connected into the same grounded hospital AC power receptacle as
Isolation Transformer.
4.9.8 PRECAUTION: When the Isolation Transformer is plugged into the mains outlet, leave at
least 2 ft clearance between the back of the AtriCure
for disconnection.
4.9.9 PRECAUTION: AtriCure
101.3 kPa and the operating condition is 69.1 kPa to 101.3 kPa
4.9.10 PRECAUTION: No modification of this equipment is allowed.
System Cart only in ventilated areas and away from
®
®
System Cart is not suitable for use in the presence of a
®
®
System Cart atmospheric storage condition is 68.6 kPa to
6 of 88
O Tanks ≤ 20 cm (8") in
2
O tanks, use caution
2
System Cart and the wall to allow access
P000880.J