5.1.2 EMC Guidance and Manufacturer's Declaration

Immunity test
150 kHz to 80 MHz
80 MHz – 2,7 GHz
Portable and mobile RF communications equipment should be used no closer to any part of the ASU System including
cables, than the recommended separation distance calculated from the equation:
���� = 6/���� × √ ����
Where:
d is the separation in meters
P is the maximum output power rating of the transmitter in watts (W) according to the Service
E is the Compliance Test Level indicated above.
Interference may occur in the vicinity of equipment marked with
the following symbol:
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
a
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the ASU System or any of its components are used
exceeds the applicable RF compliance level above, the ASU System should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating components
or the entire ASU System.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Band (MHz)
Wireless Service
380 –390
TETRA 400
GMRS 460, FRS
430 – 470
460
704 – 787
LTE Band 13, 1
GSM 800/900,
TETRA 800,
800 – 960
iDEN 820, CDMA
850,
LTE Band 5
GSM 1800; CDMA
1900;
1,700 – 1,990
GSM 1900; DECT;
LTE Band 1, 3, 4,
25; UMTS
Bluetooth, WLAN,
802.11 b/g/n, RFID
2,400 – 2,570
2450,
LTE Band 7
5,100 – 5,800
WLAN 802.11 a/n
Immunity Test Level
General
General
9 of 88
P000880.J
Compliance Test
(V/m)
Level (V/m)
< 3
3
27
28
9
28
28
28
9
< 3
3
27
28
9
28
28
28
9