Error Code - Coltene CanalPro CL Instrucciones Para El Uso

Pieza de mano eléctrica inalámbrica
Ocultar thumbs Ver también para CanalPro CL:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75

12. Error Code

If the motor handpiece stops due to an abnormality such as a malfunction, overload, break or wrong use, it
automatically checks the state of the control unit and detects the cause of the abnormality and displays an
error code on the LCD panel. If an error code is displayed, turn on the power again and check whether the
same error code is displayed. If the same error code is displayed, take action by referring to the instructions
provided in the "Check/Remedy" column in the following table.
Error code Error
During rota-
tion of the
motor hand-
piece
At the time
of Charging
Other
At the time of
calibration
Cause
Self-Check error
Malfunction of circuit
Overcurrent
The motor handpiece is
locked. (at the time of
the auto reverse mode)
Overvoltage
Malfunction of circuit
Overheating of
High load was continuous-
motor
ly applied to the motor
handpiece for a relatively
long time.
Charger failure
Malfunction of charger
Low voltage of
The voltage of batteries is
batteries
too low (The life of batter-
ies or not inserted.)
High voltage of
The voltage of batteries is
batteries
too high. (Malfunction of
circuit)
Outside the
Outstanding the range of
range of work-
working temperature or
ing temperature
break in the thermistor in
the battery section
Abnormal heat
The batteries generate
generation from
abnormal heat.
batteries
Beyond the up-
The operating life of the
per limit
motor handpiece or
contra angle has expired.
Below the lower
limit
Check/Remedy
Contact your dealer
Remove load
Contact your dealer
Leave it as it is until it
cools down.
Contact your dealer
Put the batteries into the
battery chamber, or replace
with new batteries.
Contact your dealer.
Use within the range of
working temperature.
Replace the batteries. If the
heat generates from the
new set of batteries, mal-
function of the circuit may
be suspected. Contact your
dealer.
Replace the motor
handpiece or contra angle
head.
47
Tabla de contenido
loading

Tabla de contenido