AUTO REVERSE OFF
The motor handpiece stops without reverse rotation. If you want it to rotate (forward-rotate) again, re-
press the ON/OFF Key or re-step.
CAUTION
- When the battery charge goes down, the actual load may not reach the preset torque limit value. In this
case this auto reverse function will not be activated.
- If a load is continuously applied to the motor handpiece, it may automatically stop to prevent overheat-
ing. In this case, leave the motor handpiece for a while until it cools down.
- If the ambient temperature is low, the alarm may sound when the motor handpiece rotates, however,
this is not a failure. If the alarm does not stop sounding even when the ambient temperature increases,
clean the contra angle head and perform calibration. (Refer to 6 (3) Calibration)
(6) Completion of Medical Treatment
When the treatment is completed, return the motor handpiece to the handpiece stand and hold down the
POWER key for more than one second to turn off the power.
„ Auto Power Off
If the key is not operated or approximately 10 minutes passes during rotation with no-load, the power is
automatically turned OFF to save energy and prevent improper operation. However, if load is applied dur-
ing rotation of the motor handpiece, the power is not turned OFF even during no operation.
„ LAST MEMORY
The last setting used before the power is turned OFF is memorized. The setting when the power is turned
ON next is as that immediately used before the power has been turned OFF.

6. Convenient Function

(1) Program
You can change any preset values and have them memorized to your desired settings (rotation speed,
torque limit value, gear ratio and auto reverse mode)
1. Press the PROGRAM key until it turns to the program number which you want to have memorized.
2. Adjust the rotation speed, torque limit value, gear ratio and auto reverse mode by UP/DOWN key accord-
ing to your needs.
3. Hold down the PROGRAM key for more than one second. When the alarm sounds the settings have been
memorized.
NOTICE
- The program cannot be memorized while the motor handpiece is in motion.
- The program is not memorized unless the PROGRAM key is held down for more than one second. If the
program number is changed by the PROGRAM key, the initially memorized preset values remain. (Cancel
function)
(2) Initialization of Program (Factory-set configuration)
The program can be returned to the original state configured at the time of factory shipment, if setup be-
comes confusing.
1. Turn off the power, when the power is on.
2. Hold down the PROGRAM key and ON/OFF key for more than one second, while pressing the POWER key
at the same time.
3. Release the keys when the LCD panel indicates "
played.
4. Initialization is completed, when "
42
", and re-press the PROGRAM key while this is dis-
"is displayed.