Error Messages; Cleaning, Maintenance And Repair; Maintenance And Repair Of Contra Angle - Coltene CanalPro CL2 Manual Del Usuario

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
5.0

Error Messages

Error codes are displayed on the LCD and are described below.
Errors are cleared by pressing the handpiece On/Off switch. The unit
will then return to standby mode.
Error
Code
The preset torque has
E1
been exceeded.
Battery Voltage is
E2
Excessively high.
Battery Current Error is
E3
detected.
Battery voltage too low
E4
while charging.
Battery voltage too
E5
high while charging.
Battery Temperature is
E6
excessively high
Motor Driver Circuit is
E7
Defective
6.0

Cleaning, Maintenance and Repair

6.1

Maintenance and Repair of Contra Angle

The Contra Angle is the only CanalPro
Lubrication is required prior to each use as well as before sterilization.
6.1.1 Lubricate once a day prior to sterilization. Insert the spray nozzle into the
Error
1. The handpiece is
locked.
2. The load is exceeding
the preset torque.
1. Battery type is incorrect
2. The circuit is defective.
Current is excessively
low even in the Run mode
or when the current is
excessively high during
operation.
Voltage is low within 10
minutes after charging.
Voltage is high 10 minutes
after charging.
Battery temperature
abnormally high.
Circuit board damaged.
end of the contra angle that attaches to the handpiece. Spray the lubricant
5 times.
CanalPro
Note
Cause
1. Check if the angled
2. Increase the preset torque
1. Replace the battery.
2. Contact customer service.
Contact customer service.
Replace the battery.
Replace the battery.
Contact customer service.
Replace the battery.
Contact customer service.
Contact customer service.
CL2 attachment that can be lubricated.
TM
CL2 Cordless Handpiece
TM
Solution
portion is locked.
value.
13

Solución de problemas

loading